SEMICON & HIGH PURITY WATER
The monitoring of “Ultra Pure Water” or “High Purity Water” requires the lowest possible detection limits for process analyzers. Resistivity is often around 18 MΩ. Dedicated instruments can detect infinitely small amounts (ppt e.g. parts per trillion) of ions.

Chlorine, Ozone, ClO2. The pretreatment phase uses identical analyzers for Chlorine and Ozone as potable water plants. The measurement after dechlorination is more critical because the chlorine content must be zero.
AMI Trides is the best choice for such applications. The Trides sensor has virtually no drift. The absence of sample and the failure of sensor cleaning will trigger an alarm. Hence there are no false zeros. Samples must have a conductivity superior to 5 µSiemens.
AMI Codes is the right choice for samples with a conductivity below 5 µSiemens.
CHEMATEST portable photometer: A high precision portable photometer for chlorine, ozone and other disinfectants is available for calibration and QC.

CONDUCTIVITY
Monitor AMI Rescon
Complete monitoring system for the automatic, continuous measurement of the specific resistivity / specific conductivity in high purity water.

  • Complete system mounted on stainless steel mounting panel
  • Transmitter AMI Rescon in a rugged aluminum enclosure (IP 66).
  • High precision resistivity sensor, Swansensor RC-U, a two wire electrode made of stainless steel with integrated precision temperature probe for automatic temperature compensation.
  • Stainless steel flow cell QV-Hflow with needle valve and digital sample flow measurement.
  • Automatic on-board electronic stability test with ultra high precision resistors.
  • Factory tested and calibrated on high purity water. Ready for installation and start-up.

Datasheet Monitor AMI ResconEinfluss der Temperatur auf die elektrische LeitfähigkeitInfluence of Temperature on Electrical Conductivity

QA-Monitor AMI INSPECTOR Resistivity
The portable AMI INSPECTOR is a complete monitoring system mounted on a small panel with supporting stand and a rechargeable battery for stand-alone operation, designed as an inspection equipment for quality assurance of online, resistivity process monitors.

  • Resistivity measurement range: 0.01 to 18.18 MΩ-cm
  • Rechargeable battery for stand-alone operation for at least 24 hours.

AMI Inspector Resistivity (Data sheet)
Influence of Temperature on Electrical Conductivity

pH
Monitor AMI pH-Redox
Monitor for continuous measurement of pH or redox (ORP) in high purity water, steam and condensate.

Complete system mounted on stainless steel panel:

  • Transmitter AMI pH-Redox in a rugged aluminum enclosure (IP 66).

Measuring range: 1 to 13 pH respectively -500 to +1’500mV depending on installed sensor.

  • Various combined or separated sensors with reference electrodes available.
  • Flow cell QV-Flow IS1000 made of stainless steel with quick release vessel for easy calibration, needle valve, digital sample flow meter and temperature sensor.
  • Factory tested, ready for installation and operation.
  • Special features for high purity water with Nernst plus solution temperature compensation.

Monitor AMI ph redox m flow (Data sheet)Flyer AMI Phredox QV flow

QA-Monitor AMI INSPECTOR pH

  • The portable AMI INSPECTOR is a complete monitoring system mounted on a small panel with supporting stand and a rechargeable battery for stand-alone operation, designed as an inspection equipment for quality assurance of online, pH process monitors.
  • Measurement range: 1 to 12 pH
  • Rechargeable battery for stand-alone operation for at least 24 hours.

AMI Inspector ph (Data sheet)

REDOX(ORP)
Monitor AMI PH Redox QV Flow

Monitor for continuous measurement of pH or redox (ORP) in high purity water, steam and condensate. Complete system mounted on stainless steel panel:

  • Transmitter AMI pH-Redox in a rugged aluminum enclosure (IP 66).
  • Measuring range: 1 to 13 pH respectively -500 to +1’500mV depending on installed sensor.
  • Various combined or separated sensors with reference electrodes available.
  • Flow cell QV-Flow IS1000 made of stainless steel with quick release vessel for easy calibration, needle valve, digital sample flow meter and temperature sensor.
  • Factory tested, ready for installation and operation.
  • Special features for high purity water with Nernst plus solution temperature compensation.

Monitor AMI ph redox qv flow (Data sheet)Flyer AMI Phredox QV Flow

Oxygen (dissolved)
Monitor AMI Oxytrace
High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water.

  • Range: 0-20 ppm or 0-200 % saturation (at 25°C).
  • Automatic air pressure and temperature compensation.
  • Short rinse-down time due to integrated guard.
  • Factory tested, ready for use.

Monitor AMI Oxytrace QV Flow (Data sheet)

Monitor AMI Oxytrace QED
High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water. Integrated auto verification with faraday electrode setup.

  • Range: 0-20 ppm or 0-200 % saturation (at 25°C).
  • Automatic air pressure and temperature compensation.
  • Short rinse-down time due to integrated guard.
  • Faraday electrode and control unit for the automatic or manual verification.
  • Factory tested, ready for use.

Monitor AMI Oxytrace QED (Data sheet)Flyer AMI Oxytrace QED

Monitor AMI Oxytrace QED; compact version
High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water.
Integrated auto verification with faraday electrode setup.

  • Mounted on a small panel (width: 400mm, height 420mm)
  • Range: 0-20 ppm or 0-200 % saturation (at 25°C).
  • Automatic air pressure and temperature compensation.
  • Short rinse-down time due to integrated guard.
  • Faraday electrode and control unit for the automatic or manual verification.
  • Factory tested, ready for use.

Monitor AMI oxytrace QED (Data sheet)

QA-Monitor AMI INSPECTOR Oxygen
The portable AMI INSPECTOR is a complete monitoring system mounted on a small panel with supporting stand and a rechargeable battery for stand-alone operation, designed as an inspection equipment for quality assurance of online, trace oxygen process monitors.

  • Measurement range:
  • 0.01ppb to 20ppm O2 (at 25°C) or
  • 0 to 200% saturation
  • Rechargeable battery for stand-alone operation for at least 24 hours

AMI Inspector Oxygen (Data sheet)

Monitor AMI Oxytrace; compact version
High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water.

  • Mounted on a small panel (width: 280 mm, height: 180 mm)
  • Range: 0-20 ppm or 0-200 % saturation (at 25°C).
  • Automatic air pressure and temperature compensation.
  • Short rinse-down time due to integrated guard.
  • Factory tested, ready for use

Monitor AMI Oxytrace QV Flow (Data sheet)

SILICA
Analyzer AMI Silitrace Ultra
Complete monitoring system for the automatic, continuous measurement of trace amounts of silica in ultra pure water.

  • Measuring range: 5 ppt (0.005 ppb) to 25 ppb
  • Based on colorimetric measurement principle
  • Complete system including: measurement and control electronics, photometer with integrated temperature controlled reaction chamber, flow indicator, reagent dosing system and reagent containers using reversed osmosis system to concentrate the sample (Carrcentrator)
  • Optional communication interface

Monitor AMI Silitrace Ultra (Data sheet)

Analyzer AMI Silitrace
Complete monitoring system for the automatic, continuous measurement of silica in water steam cycles and high purity water.

  • Measuring range: 0.5 to 1’000 ppb
  • Based on colorimetric measurement principle
  • Complete system including: measurement and control electronics, photometer with integrated temperature controlled reaction chamber, flow indicator, reagent dosing system and reagent containers.
  • Optional communication interface
  • Sample Sequencer for up to 6 sample streams as an accessory

Monitor AMI Silitrace (Data sheet)

SODIUM
Analyzer AMI Sodium A
Sodium analyzer for steam, condensate and high purity water in the range 0.1 – 10’000 ppb.

  • For acidic samples with low pH ≥ 2 (e.g. sampling after cation exchangers).
  • Active alkalization reagent addition system for diisopropylamine with continuous pH monitoring.
  • Simple two-point calibration with storage of calibration history.
  • Easy to use grab sample capability.
  • Available in a single or dual stream version.

AMI Sodium A (Data sheet)Flyer AMI Sodium A

Analyzer AMI Sodium P; 28cm
Sodium analyzer for steam, condensate and high purity water in the range 0.1 – 10’000 ppb.

  • For samples with pH ≥7
  • Reliable alkalization reagent addition system for diisopropylamine or ammonia with continuous pH monitoring.
  • Simple two-point calibration with storage of calibration history.
  • Easy to use grab sample capability.
  • Dual stream version as an option available.

AMI Sodium P (Data sheet)Flyer AMI SodiumP Low Water

Analyzer AMI Soditrace
Analyzer for the continuous measurement of sodium ions in trace amounts in high purity water applications and steam generation.

  • Complete Sodium analyzer panel-mounted for easy wall installation.
  • Lowest available detection limit for sodium ion concentrations of 0.001 ppb.
  • Programmable automatic calibration by multiple known standard addition at low levels.
  • Programmable automatic regeneration of sodium electrode (pat.pending).
  • Automatic control of sample pH conditioning.
  • Automatic temperaure compensation.
  • Continuous monitoring of reagent exhaustion, sample flow and sample temperature.

AMI Soditrace (Data sheet)Flyer AMI Soditrace

TOC
On-line analyzer for the continuous measurement of Total Organic Carbon in pure and ultrapure water.

  • Reagent-free measurement of TOC by UV oxidation and differential conductivity detection.
  • System suitability test according USP <643> and EP 2.2.44.
  • Automatic addition of standard solutions.
  • Automatic dilution of standards to 500 ppb TOC as Sucrose and to 500 ppb TOC as Benzoquinone.
  • Reaction Time: < 2 Minute
  • Factory tested and ready for installation and operation

Monitor AMI Linetoc (Data sheet)Flyer AMI LineTOCToc Messung In Reinstwasser

TURBIDITY
Monitor AMI Turbitrace
High Precision Nephelometer complying with ISO 7027 (EN 27027, DIN 38404) for the continuous measurement of low level turbidity in pure and ultra pure water.

  • Range: 0.000 to 100.0 FNU with a precision of ± 0.001 FNU or 1% of reading.
  • Programmable automatic zero point measurement for drift compensation by integrated sub-micron particle filter.
  • Monitoring of sample flow and sensor fouling with alarms.
  • High precision secondary standards for quality assurance

AMI Turbitrace (Data sheet)Flyer Turbitrace

Monitor AMI Turbitrack
High Precision Nephelometer complying with ISO 7027 (EN 27027, DIN 38404) for the continuous measurement of low level turbidity in pure water with up to 10 bar.

  • Range: 0.000 to 100.0 FNU with a precision of ± 0.001 FNU or 1% of reading.
  • Turbidimeter with sample chamber purge and flow controller.
  • Monitoring of sample flow with alarm.
  • High precision secondary standards for quality assurance.

AMI Turbitrack (Data sheet)Flyer Turbitrack